Akbarnia, B. A., K. Cheung, H. Noordeen, H. Elsebaie, M. Yazici, Z. Dannawi, and N. Kabirian, "Next generation of growth-sparing techniques: preliminary clinical results of a magnetically controlled growing rod in 14 patients with early-onset scoliosis.", Spine, vol. 38, issue 8, pp. 665-70, 2013 Apr 15. Abstract

STUDY DESIGN: Prospective nonrandomized study.

OBJECTIVE: To report the preliminary results of magnetically controlled growing rod (MCGR) technique in children with progressive early-onset scoliosis.

SUMMARY OF BACKGROUND DATA: The growing rod (GR) technique is a viable alternative for treatment of early-onset scoliosis. High complication rate is attributed to frequent surgical lengthening. The safety and efficacy of MCGR were recently reported in a porcine model.

METHODS: Multicenter study of clinical and radiographical data of patients who underwent MCGR surgery and at least 3 distractions. Distractions were performed in clinic without anesthesia/analgesics. T1-T12 and T1-S1 heights and the distraction distance inside the actuator were measured after lengthening.

RESULTS: Fourteen patients (7 girls, 7 boys) with a mean age of 8 years, 10 months (3 yr, 6 mo to 12 yr, 7 mo) had 14 index surgical procedures. Of the 14, 5 had single-rod (SR) surgery and 9 had dual-rod (DR) surgery, with overall 68 distractions. Diagnoses were idiopathic (N = 5), neuromuscular (N = 4), congenital (N = 2), syndromic (N = 2), and neurofibromatosis (N = 1). Mean follow-up was 10 months (5.8-18.2). The Cobb angle changed from 60° to 34° after initial surgery and 31° at latest follow-up. During distraction period, T1-T12 height increased by 7.6 mm for SR (1.09 mm/mo) and 12.12 mm for DR (1.97 mm/mo). T1-S1 height gain was 9.1 mm for SR (1.27 mm/mo) and 20.3 mm for DR (3.09 mm/mo). Complications included superficial infection in 1 SR, prominent implant in 1 DR, and minimal loss of initial distraction in 3 SR after index. Partial distraction loss observed after 14 of the 68 distractions (1 DR and 13 SR) but regained in subsequent distractions. There was no neurological deficit or implant failure.

CONCLUSION: Preliminary results indicated MCGR was safe and provided adequate distraction similar to standard GR. DR achieved better initial curve correction and greater spinal height during distraction compared with SR. No major complications were observed during the follow-up.

Dannawi, Z., F. Altaf, N. S. Harshavardhana, H. El Sebaie, and H. Noordeen, "Early results of a remotely-operated magnetic growth rod in early-onset scoliosis.", The bone & joint journal, vol. 95-B, issue 1, pp. 75-80, 2013 Jan. Abstract

Conventional growing rods are the most commonly used distraction-based devices in the treatment of progressive early-onset scoliosis. This technique requires repeated lengthenings with the patient anaesthetised in the operating theatre. We describe the outcomes and complications of using a non-invasive magnetically controlled growing rod (MCGR) in children with early-onset scoliosis. Lengthening is performed on an outpatient basis using an external remote control with the patient awake.Between November 2009 and March 2011, 34 children with a mean age of eight years (5 to 12) underwent treatment. The mean length of follow-up was 15 months (12 to 18). In total, 22 children were treated with dual rod constructs and 12 with a single rod. The mean number of distractions per patient was 4.8 (3 to 6). The mean pre-operative Cobb angle was 69° (46° to 108°); this was corrected to a mean 47° (28° to 91°) post-operatively. The mean Cobb angle at final review was 41° (27° to 86°). The mean pre-operative distance from T1 to S1 was 304 mm (243 to 380) and increased to 335 mm (253 to 400) in the immediate post-operative period. At final review the mean distance from T1 to S1 had increased to 348 mm (260 to 420).Two patients developed a superficial wound infection and a further two patients in the single rod group developed a loss of distraction. In the dual rod group, one patient had pull-out of a hook and one developed prominent metalwork. Two patients had a rod breakage; one patient in the single rod group and one patient in the dual rod group. Our early results show that the MCGR is safe and effective in the treatment of progressive early-onset scoliosis with the avoidance of repeated surgical lengthenings.